Complex applications, such as nucleic acid purification or solid phase extraction should not tie laboratory personnel to the workbench.
The epMotion 5075v allows true walk-away automation of purification protocols that rely on vacuum filtration. The 5075v is controlled by MultiCon industry PC and can be configured with CleanCap and up to 3 thermal modules. More information

Order Info

Product Information

The epMotion 5075v is all about productivity. With its integrated vacuum station the epMotion 5075v allows true walk-away automation of purification protocols that rely on vacuum filtration.

Integrated Vacuum Station


Loaded and unloaded with the gripper, the vacuum station adapts automatically to any filter plate type. Silent operation of internal pump with no need for extra tubes, wiring, or reservoir maintenance. The unique Eppendorf silicon mat accessory helps to save time and money
by allowing you to use only portions of a 96-well plate when working with smaller sample numbers. The VAC thermo lid accessory also reduces the extraction time substantially
by accelerating the ethanol evaporation step.

epMotion® 5075v

Plug'n'Prep® downloads


The epMotion Plug'n'Prep® concept offers you freedom of choice: Purify nucleic acids with protocols from your favorite kit vendor. All methods are pre-tested by the kit supplier and ready to use on the epMotion. In order to run your Plug'n'Prep® method download the Plug'n'Prep® method- and labware file from the VIP page and transfer to your epMotion. With epMotion 5075's unique Plug’n’Prep® applications, you are free to choose virtually any kit for your purification needs and you are free from the bench with true walk-away automation. Plug’n’Prep® also means it is easy to set up—you can start running your protocols on the day of installation!
More customized applications can be developed easily using the epBlue software.

epMotion® 5075v

Gripper Tool


The gripper tool allows transport of labware on the worktable of epMotion 5073 or 5075 versions. It is loaded by automatic tool exchange and can be programmed easily in the software. The stacking of up to 5 plates is possible.

epMotion® 5075v

Features

  • Same as epMotion 5075l
  • 12 SLAS/ANSI deck positions
  • Integrated vacuum pump – silent operation, no tubing, wiring, or reservoirs to maintain
  • Vacuum station fully integrated that automatically adapts to filter plates controlled by software
  • Gripper for plate and labware transport
  • Three thermal module options (2 in case of 5075vt)
  • Also available as 5075vt version with integrated Eppendorf ThermoMixer®

Applications

  • Like 5075l with vacuum purification technologies
  • Solid phase extraction
  • Filtration
  • DNA from blood, tissue and cells
  • DNA from plants and bacteria
  • Sequencing and PCR clean-up
  • Total RNA isolation
 epMotion® 5075v
Gripper carrying capacity1,200 g
Operationelectronic
Pipetting typeair-cushion
SLAS/ANSI deck positions12
Volume range0.2 µL – 1 mL
Power supply100 – 240 V ±10 %, 50 – 60 Hz ±5 %
Max. power consumption700 W
Dimensions (W × D × H)107 × 61 × 67 cm / 43 × 24 × 27 in
Weight w/o accessories86 kg / 189.6 lb
Thermal module (optional) 
Cooling rate4 °C/min
Heating rate9 °C/min
Temperature range0 °C – 110 °C
Vacuum unit 
Suction range0.1-85 kPa
Suction time1 sec to 36 min
Vacuum max. output35 NL/min
Conductor – X, Y, Z Positioning 
Random measurement error±0.1 mm
Systematic measurement error±0.3 mm
Detector 
Optical confocal infrared detectorContact-free recognition of fill level, inserted tools, labware, tip types and quantities
Optical sensorContact-free recognition of fill level, inserted tools, labware, tip types and quantities
Dispensing tool 
Random measurement error (1 µL)1,2≤5 %
Random measurement error (1,000 µL)1,2≤0.15 %
Random measurement error (50 µL)1,2≤0.4 %
Systematic measurement error (1 µL)1,2±15 %
Systematic measurement error (1,000 µL)1,2±0.7 %
Systematic measurement error (50 µL)1,2±1.2 %
1in pipetting mode, free jet, without pre-wetting, with distilled water, at 20 °C
2See Eppendorf Application Note 168 for typical pipetting performance

You might also consider